The following procedure describes how you can quantify a compound with a Manual Factor instead of using a calibration curve.
Prerequisites
To be able to carry out the procedure as described, you need the privilege Processing Method > Edit calibration parameters. Privileges are configured in the Control Panel.
If the project enforces method approval, the method status must be Generic. For details, see Method approval.
If the Processing Method window is hidden, click Processing Method in the Home ribbon tab to show the window.
In the Processing Method window, select the Compounds > Calibration node.
Under Mode, select Manual Factor.
The Manual Factor column becomes editable.
Under Manual Factor, enter the response factor you want to use.