Manual integration corrections may be saved in either
a processing method, or
a specific injection.
If they are saved in a processing method, they will be applied to any data that is reprocessed with this method. To avoid that they will be applied again, remove them from the processing method.
To remove manual integration corrections from a specific injection, reprocess the injection with a suitable processing method using the Clear corrections option.
In the processing method, select the Integration Events > Manual Integration node.
Right-click the table, and select the required command from the context menu. You can delete all manual integration corrections from all signals, or only the selected manual integration corrections.
If you do not delete saved manual integration corrections, they will be applied again during the next reprocessing.
Make sure that the method assigned to the injection does not contain manual integration corrections.
In the Processing ribbon tab, open the submenu of the Reprocess button, then select Reprocess - Clear corrections.
Reprocessing with the Clear corrections option removes any manual integration corrections from the signal.