Order number |
Order number of compound (as sorted by retention time). The order number cannot be edited. | |||
Type |
The icon indicates whether it is an internal or external standard, a normal or time reference compound, or a group compound.
External standard, normal compound
External standard, time reference compound
Internal standard, normal or time reference compound
Named group
Timed group | |||
Name |
The name of the compound. | |||
Signal |
The detector signal used to identify and quantify the compound. The drop-down list offers the names of all signals that are currently shown in the Method Selector or have been used in the method during the current session. The list of signal names is independent of the instrument driver.
| |||
Impurity sensitivity [%] | By changing the sensitivity value (default: 50 %), you influence the threshold value and thus the criteria for a green or red coloring of the UV Purity value in the injection results. To check for UV impurities within a peak, you can compare all spectra recorded for a specific peak with the spectrum at the peak apex. Here, you set the parameters used to calculate the UV impurities. | |||
UV reference spectrum | The icon
To confirm the compound identity, you can compare a UV reference spectrum with the current spectrum of a peak. Here, you set the parameters used to calculate UV confirmation match factor and to display it in the injection results. | |||
MS reference spectrum | The icon
|