If an analysis show no chromatographic peaks, only a flat baseline or minor noise, run autotune. If the MSD passes, the problem is most likely related to the GC. Check the possible causes listed under Passes tune. If the MSD does not pass, the problem is most likely in the MSD. Check the possible causes listed under Does not pass tune.