Only MSDs equipped with a diffusion pump have a foreline pressure gauge.
On MSDs equipped with a diffusion pump, column (carrier gas) flow should not exceed 2.0 mL/minute. On MSDs equipped with a turbomolecular pump, column (carrier gas) flow should not exceed 4.0 mL/minute. Reduce the column flow and retune the MSD.
Replace the foreline gauge. See the MSD maintenance documentation for instructions for replacing the foreline gauge.
Replace the foreline pump or contact your Agilent Technologies service representative.