It normally takes the diffusion pump 10 - 15 minutes to reach its normal operating temperature. It may take slightly longer if:
The laboratory is cooler than normal
The fluid level in diffusion pump is especially high
The foreline pump is not working as well as normal
Column flow is higher than normal
It normally takes the turbomolecular pump 5 - 10 minute to reach at least 80% of its normal operating speed. It may take slightly longer if:
The foreline pump is not working as well as normal
Column flow is higher than normal
Replace the foreline pump or contact your Agilent Technologies service representative.
If the level of the fluid in the diffusion pump is too high, it can slow the pumpdown process or prevent the diffusion pump from reaching its normal operating temperature. Check the level of the fluid in the diffusion pump. Refer to the MSD maintenance documentation for instructions for checking the fluid level in the diffusion pump.
If the components of the stack in the diffusion pump are not correctly aligned, it can prevent the diffusion pump from reaching its normal operating temperature. The stack is not removed for normal maintenance so this should be fairly uncommon.
Contact your Agilent Technologies service representative.