Path/Column (Valve Thermostat Cluster)

Click the down-arrow to display a table containing details of all the configured paths, including the columns, with important details about the column parameters. Select the path/column to use for the method from the table.

Click this button to load the plumbing configuration currently stored in the Valve Thermostat Cluster. The updated information is displayed in the visual representation of the Valve Thermostat Cluster.

Visualization

The graphic gives a visual representation of the Valve Thermostat Cluster configuration. If a Multicolumn Thermostat is equipped with a Thermal Equilibration Device, the columns are partially covered with a dark triangle.

Place the mouse cursor on a column to display a tooltip containing the column tag information. Clicking on a column highlights the corresponding line(s) in the Temperature Settings table in the Advanced section.

Enforce Column for Run

Mark this check box to restrict the execution of the method to a specific column type, identified by its product number. Either enter the product number or select it from the drop down list, which contains all columns that have been defined in the Column Assignment dialog box and contain a product number. Note that product numbers in the Column Assignment dialog box are hidden by default, and can be displayed by clicking >> for the extended table view. Note also that the valve position selected in the Plumbing section of the Column Assignment dialog box still needs to match.