Change PMT Gain
(FLD Timetable)
PMT Gain is the
photomultiplier gain. Each step approximately doubles the signal. A typical
setting is between 10 and 15; use lower settings for very high concentrations.
Higher PMT Gain may improve signal-to-noise.
Limits: 0 to 18 in steps of 1. Default value is
10