Start Rinse (Fraction Collector II Timetable)

This command allows you to rinse the fraction collector; in the case of a cluster, each fraction collector is rinsed sequentially.

The procedure starts with the master module.

The procedure rinses all fraction collectors in sequence until

  • either all fraction and recovery collectors have been rinsed,

  • or the next entry in the timetable is encountered.

The duration of each rinse is calculated internally based on the flow of the linked pump and the number of repetitions defined in the Rinse dialog box.

After the rinse is complete, the collector(s) use the fraction mode Off until a new Change Fraction Mode command is encountered in the timetable. If a new fraction mode is specified after the rinse procedure in the timetable, fraction collection resumes at the fraction collector that was being used when rinsing was started.