Intended Pump Flow allows you to specify the pump flows that you expect during a run.
This information is used to calculate the expected Split ratio and Dilution factor and help you choose the appropriate Split Mode for your application.
Main Flow |
Enter the intended flow (in mL/min) of your main pump that delivers your main (preparative) flow. |
Makeup Flow |
Enter the intended flow of your makeup pump that delivers a reduced flow, for example as makeup flow to an MSD or auxiliary detector. |
The values you enter here are for preview purposes only. They are not used to control the pumps. Ensure that you set the pump flows in the individual pump method parameters accordingly. |