1.19 Instrument cannot achieve proper rough pump vacuum in a fixed amount of time. Vent and power cycle is needed to clear condition.

Probable Causes

Big leak in the vacuum system

No oil in rough pump

Oil mist filter is full of oil

Bad rough pump

Suggested Actions

Check that rough pump is plugged in.

Check rough pump oil level and ballast.

Check for leaks (block the capillary with septa).

Check all vacuum clamps and hoses.

Check other vacuum seals such as the top plate.

Fill in oil in rough pump.

Change oil mist filter.

Change rough pump.

If this error continues to occur, copy this error on the screen and contact Agilent for more help.