These messages apply to the LC/MSD and LC/MSD XT Single Quadrupole instruments.
1.0 Fragmentor driver cannot maintain the requested voltage.
1.3 Octopole shroud driver cannot maintain the requested voltage.
1.12 Lens2 RF driver cannot maintain the requested RF peak voltage.
1.13 Octopole RF driver cannot maintain the requested RF peak voltage.
1.15 MS U- driver cannot maintain the requested mass setting.
1.16 MS U+ driver cannot maintain the requested mass setting.
1.17 MS RF driver cannot maintain the requested mass setting.
1.18 MS U+/U- DC-to-DC converter supply is not producing voltage or is over producing voltage.
2.0 Main board lens driver power supply cannot maintain the voltage.
2.16 Firmware timing problem: log amp is read before conversion is completed.
2.20 AC line zero cross circuit has failed. Vent and power cycle is needed to clear condition.
2.26 MS heater has exceeded its maximum temperature. Power cycle is needed to clear condition.