2.19 Selector stream valve cannot switch properly.

Probable Causes

Stream valve is bad

Bad valve PCA

Suggested Actions

Replace stream valve.

If this error continues to occur, copy this error on the screen and contact Agilent for more help.

See Also

Error Messages (LC/MS)

Tune the instrument