2.26 MS heater has exceeded its maximum temperature. Power cycle is needed to clear condition.

Probable Causes

Heater thermal runaway

Temperature above highest setpoint

Suggested Actions

Power cycle the instrument.

Reduce heater temperature.

If this error continues to occur, copy this error on the screen and contact Agilent for more help.

See Also

Error Messages (LC/MS)

Tune the instrument