Advanced Acquire > Ramp Table (LC/MS)

You only have one fragmentor ramp, and it applies to all scans that enable the Fragmentor ramp.

The method audit trail notes any ramp table change for every ramped scan to which that ramp applies.

What do you want to do?

Edit the Ramp Table

Edit the Acquisition Parameters Table

Edit Advanced Acquire section

Edit a method

Ramp pop-up window

Select the check box in the Fragmentor ramp? column in the Advanced Acquire > Acquisition Section to ramp the Fragmentor during acquisition. Click the arrow in that column to open the Ramp Table. This pop-up window shows a table containing pairs of Mass (m/z) and Fragmentor voltage (V). The value for the fragmentor is ramped between the values in this table.

All Scan rows in the table that have Positive polarity use one Ramp Table.

All Scan rows in the table that have Negative polarity use a different Ramp Table.

SIM rows do not use the Ramp Table; you specify the Fragmentor (V) to use.

Ramp table

These two columns are in the Ramp Box: Mass (m/z) and Fragmentor voltage (V).

The table automatically updates so that the masses in the Mass (m/z) column are in increasing order.

The fragmentor voltage for each mass acquired is determined by ramping the fragmentor between the two nearest Mass (m/z) values.