The OPA/FMOC-protocol for AAA workflow allows you to perform the derivatization of the amino acids directly in the needle/loop prior to injection. This procedure is described in the Application Note 5994-2189.
The sample is taken from the position specified in the sample information or sequence/worklist. The mixture is prepared within the needle and directly injected. For more details, mark the Advanced check box at the top right of the workflow view to display the injector program in the advanced view.
The following parameter settings are available:
Parameter | Description |
---|---|
Sample Volume | Specify the sample volume. |
Borate Buffer Volume | Specify the volume of the borate buffer. |
From Location | Enter the location of the borate buffer. See Specifying a Location. |
OPA Reagent Volume | Specify the volume of the OPA reagent. |
From Location | Enter the location of the OPA reagent. See Specifying a Location. |
Mix 1 Volume | Specify the air volume for the mixing within the needle. |
Repetitions | Select the number of repeat cycles for the mixing. |
| |
FMOC Reagent Volume | Specify the volume of the FMOC reagent. |
From Location | Enter the location of the FMOC reagent. See Specifying a Location. |
Mix 2 Volume | Specify the air volume for the second mixing within the needle. |
Repetitions | Select the number of repeat cycles for the mixing. |
| |
Injection Diluent Volume | Specify the volume of the injection diluent. |
From Location | Enter the location of the injection diluent. See Specifying a Location. |
Mix 3 Volume | Specify the air volume for the third mixing within the needle. |
Repetitions | Select the number of repeat cycles for the mixing. |
| |
Accumulated Draw Volume | The sum of all entered volumes plus 0.7 µL per standard wash cycle. The maximum draw volume is limited by the loop volume and the syringe volume. If the accumulated draw volume exceeds the maximum draw volume, a warning icon is displayed. |