Enables the autosampler to draw from a specified source.
Parameter | Value | Description |
---|---|---|
Draw | Default | The default amount specified in the Set Up Standard Autosampler panel. |
| Maximum | The maximum amount defined by the injector configuration. |
| <value> (µL) | An amount that overrides the default injection volume. |
from | Sample | From the location specified in the sample information or sequence/worklist. |
| Location | From a specified location. |
| Seat | From the seat capillary. |
| Air | From air, above the seat capillary. |
| Vial+ | Vial+ defines an offset from the current sample vial number. For example, if the current sample vial is 1, a value for Vial+ of 3 specifies a vial that is three positions away from the sample vial, in this case vial 4. The limits depend on the sample tray configuration: -15 to +15 for the 8 ml-vial tray -40 to +40 for the 2 ml-vial half-tray -100 to +100 for the 2 ml-vial full tray
|
Speed | Default | The default speed, specified in the Set Up Standard Autosampler panel. |
| Maximum | The maximum draw speed |
| <value> (µL/min) | A value that overrides the default draw speed. |
useOffset | Default | The default draw position, specified in the Set Up Standard Autosampler panel. |
| <value> (mm) | -2.5 to +35 mm below or above the standard draw position The offset is calculated from the bottom of the vial, with a slight amount of clearance to cover tolerances. Use a negative offset to touch the vial bottom with the needle. Use a positive offset to raise the needle higher in the vial to clear precipitates on the bottom of the vial, or to select one phase of a biphasic sample. |
The parameters shown are dependent on the selected source.