Moves to a specified position and ejects a specified amount from the needle.
Parameter | Value | Description |
---|---|---|
Eject | Default | The default amount specified in the Set Up Standard Autosampler panel. |
| Maximum | The maximum amount defined by the injector configuration. |
| <value> (µL) | A specified amount that overrides the default amount. |
to | Sample | The location specified in the sample information or sequence/worklist. |
| Location | A specified location. |
| Seat | The seat capillary. |
| Vial+ | Vial+ defines an offset from the current sample vial number. For example, if the current sample vial is 1, a value for Vial+ of 3 specifies a vial that is three positions away from the sample vial, in this case vial 4. The limits depend on the sample tray configuration: -15 to +15 for the 8 ml-vial tray -40 to +40 for the 2 ml-vial half-tray -100 to +100 for the 2 ml-vial full tray
|
Speed | Default | The default speed specified in the Set Up Standard Autosampler panel. |
| Maximum | The maximum eject speed. |
| <value> (µL/min) | A value that overrides the default eject speed. |
useOffset | Default | The default offset position, specified in the Set Up Standard Autosampler panel. |
| <value> (mm) | -2.5 to +35 mm below or above the normal offset position |
The parameters shown are dependent on the selected source.