Moves the autosampler needle as specified.
Parameter | Value | Description |
---|---|---|
Up |
| Moves the needle to its upper position. |
Down |
| Moves the needle to its lower position. |
Into |
| Moves the needle into a specified Source with a specified Offset.The Source can be Sample, Location or Vial+. |
| Sample | The location specified in the sample information or sequence/worklist with a specified Offset. |
| Location | A specified location. |
| Vial+ | Vial+ defines an offset from the current sample vial number. For example, if the current sample vial is 1, a value for Vial+ of 3 specifies a vial that is three positions away from the sample vial, in this case vial 4. The limits depend on the sample tray configuration: -15 to +15 for the 8 ml-vial tray -40 to +40 for the 2 ml-vial half-tray -100 to +100 for the 2 ml-vial full tray
|
Offset (mm) | Def. | The default draw position, specified in the Setup Injector dialog box. |
| Value | -2.5 to +35 mm below or above the normal draw position The offset is calculated from the bottom of the vial, with a slight amount of clearance to cover tolerances. Use a negative offset to touch the vial bottom with the needle. Use a positive offset to raise the needle higher in the vial to clear precipitates on the bottom of the vial, or to select one phase of a biphasic sample. |
The parameters shown are dependent on the selected source.