Moves the valve to a specified position
Parameter | Value | Description |
---|---|---|
Mainpass |
| Turns the valve to the main flow path position |
Main+Start |
| Turns the valve to the main flow path position and generates a start pulse.
|
Bypass |
| Turns the valve to the bypass position. |
Main In |
| Turns the valve to the main flow path position while the needle is at the Source for a Time between 1 and 99 seconds. The source (from can be a Sample, a Location, Seat, the wash port, Sample+ or Vial+. |
| Sample | The position specified in the sample information or sequence/worklist. |
| Location | A specified location. The location can be a well or a vial. |
| Seat | The seat capillary. |
| Washport | The wash port. |
| Sample+ | Sample+ is used for wellplate sample locations only; it defines an offset from the current location. You specify the offsets for Tray, Plate, Row and Column individually. For example, if the current location is P1-A-01, values for Plate, Row and Column of 1 define location P(1+1)-(A+1)-(01+1), which is P2-B-02. The limits are: Tray: -1 to +1 (only valid if two trays are inserted) Plate: -2 to +2 (only valid for wellplates) Row: -15 to +15 Column: -23 to +23 |
| Vial+ | This function is valid for vial trays only. Vial+ is used for sample vial locations only; it defines an offset from the current sample vial number. For example, if the current sample vial is 1, a value for Vial+ of 3 specifies a vial that is three positions away from the sample vial, in this case vial 4. The limits are independent of the sample tray configuration: -99 to +99 |
|
| to (only for Sample, Location, Sample+ and Vial+) is the time to leave the valve in the main flow path (in seconds, range is 0 – 9999) use Offset (only for Sample, Location, Sample+ and Vial+) is the vertical offset of the needle relative to the default position. Range is Default as defined in Draw Position in the method, or -10 – 50. The offset of the autosampler needle is calculated from the bottom of the well, with a slight amount of clearance to cover tolerances. Use a negative offset to touch the well bottom with the needle. Use a positive offset to raise the needle higher in the well. |