Moves the valve to a specified position
Parameter | Value | Description |
---|---|---|
Mainpass |
| Turns the valve to the main flow path position. |
Main + Start |
| Turns the valve to the main flow path position and generates a start pulse.
|
Bypass |
| Turns the valve to the bypass position. |
Main In |
| Turns the valve to the main flow path position while the needle is at the source for a time between 1 and 99 seconds. The source can be a Sample, a Location, Seat, Air, Sample+ or Location+. |
| Sample | The position specified in the sample information or sequence/worklist. |
| Location | A specified location. |
| Seat | The seat capillary. |
| Air | Air, above the seat capillary. |
| Sample+ | A specified offset from the current location. Sample+ is intended for trays that support linear addressing only, for example, Vial 1, Vial 2 .... |
| Location+ | A specified offset from the current location. Location+ is intended for trays that support row and column addressing only. |