Moves the Vialsampler needle as specified.
Parameter | Value | Description |
---|---|---|
Up |
| Moves the needle to its upper position. |
Seat |
| Moves the needle to the seat. |
Into |
| Moves the needle into a specified source with a specified offset. The source can be Sample, Location, Sample+ or Location+. |
| Sample | The position specified in the sample information or sequence/worklist, with a specified Offset. |
| Location | A specified location, with a specified Offset. |
| Sample+ | A specified offset from the current location. Sample+ is intended for trays that support linear addressing only, for example, Vial 1, Vial 2 ....
|
| Location+ | A specified offset from the current location. Location+ is intended for trays that support row and column addressing only.
|
Offset (mm) | Default | The default draw position, specified in the Set Up Vialsampler dialog box. |
| Value | -2.5 to +35 mm below or above the normal draw position The offset is calculated from the bottom of the vial, with a slight amount of clearance to cover tolerances. Use a negative offset to touch the vial bottom with the needle. Use a positive offset to raise the needle higher in the vial to clear precipitates on the bottom of the vial, or to select one phase of a biphasic sample. |