The Dilution Series from Stock workflow allows you to generate a dilution series from a concentrated sample stock solution before injection. You specify a dilution factor by which all dilution steps are calculated. The sampler pipettes then the series, starting from the specified sample location and dilution solvent location, into the target locations. The specified mixing method is used to mix the diluted samples in the target sample vial.
This workflow is similar to the Dilution Series workflow, but it takes each dilution step from the sample stock solution. If you want to use a dilution series workflow that takes each dilution step from the previous dilution, use the Dilution Series workflow.
The Draw Sample parameter specifies the location of the sample.
Parameter | Value | Description |
---|---|---|
From Location | Sample | The sample is taken from the position specified in the sample information or sequence/worklist. |
Location | Enter the location in the adjacent field. See Specifying a Location. |
The Draw Dilution Solvent parameters specify the location of the dilution solvent and the dilution factor.
Parameter | Description |
---|---|
From Location | Enter the location of the dilution solvent. See Specifying a Location. |
Dilution Factor | Select the dilution factor. |
The Dilute parameters specify the target and the mixing of the diluted sample.
Parameter | Value | Description |
---|---|---|
Target Location | Enter the target locations. See Specifying a Location. Click Add to add a location or Remove to remove the selected entry. | |
Stock Volume | The resulting sample volume in the target location. | |
Diluent Volume | The resulting diluent volume in the target. | |
Target Volume | Specify the target volume. | |
Mix Mode | Select the mixing method that is used to mix the diluted sample in the target location. | |
Air bubble | The air is taken from a neutral position (e.g. wash port) and bubbled into the diluted sample. This method should be the default, as it is the most effective. For some samples, especially those containing detergents or proteins, the method may cause the sample to foam. When the sample is drawn up later, the foam may cause insufficient sample volume to be drawn up. In this case, use the other method. | |
Draw/inject from/to sample | The maximum volume of the diluted sample is drawn from and ejected in the target location. | |
Repetitions | Select the number of repeat cycles for the mixing method. | |
Speed | Specify the eject speed for the mixing with mix mode Draw/inject from/to sample. | |
Default | The eject speed specified in the Set Up panel. | |
Maximum | The maximum speed that the sampler can handle. | |
<value> (µL/min) | A value that overrides the default eject speed. |