Select a bracketing mode
Select the type of bracketing you want to use from the Bracketing
mode list.
- None: No bracketing is
applied and calibration standards and injections are processed in
order of acquisition.
- Overall: The calibration
curve is calculated with all calibration standards in the sequence,
starting with the first one and finishing with the last one. Any samples
are reprocessed after the calibration curve has been calculated.
- Non-overlap: You must have
at least three groups of standards in your sequence and at least two
standards in the middle blocks. The standards from the group in the
middle of the sequence are used in one calibration curve only.
- Overlap: You must have
at least three groups of standards in your sequence. Standards from
the group in the middle of the sequence are used in two calibration
curves (with the preceding and following blocks).