Configure dynamic ramps

  1. Request tune control and select Manual Tune > Parameters.
  2. In the Dynamic Ramps area, select an analyzer component to ramp from the menu.
  3. Select Enable this lens to allow dynamic ramping of voltages at each ion shown for this analyzer component.
  4. Specify analyzer component voltages for adjusting the mass abundance at each of the specified ions.
  5. Do a scan acquisition to view the effects of the dynamic ramping. Repeat this adjustment until you obtain the desired result.
  6. Click Save tune parameters with new tune file names and name this manual tune file.

  7. Optionally, open a pdf of the manual tune file parameters.

  8. Release tune control.