Switch to
the spare filament and when prompted to autotune click YES.
If the MSD passes tune, use this filament. Replace the first filament
when convenient.
If the second filament does not pass tune, vent the MSD and check the filament wiring connections.
Check the filaments. Replace them if necessary.
See the MSD maintenance documentation for instructions for replacing a filament.
Contact your Agilent Technologies service representative.