Sampling Table (2D-LC)
The
Sampling Table shows the 2D-LC
sampling timetable in three columns.
The title bar shows the
number of events in the Sampling
Table and the number of available events remaining. Each additional
gradient point (2D Gradient, Shifted Gradient or
Flush Gradient) reduces the total number of events available in the
Sampling Table.
Time |
The start time of the
cut. |
Function |
The mode of sampling,
Time-based Heart cut,
Time-based Comprehensive,
Start Peak-based or
End Peak-based. Click the down-arrow
to select an alternative mode. | | Start Peak-based peak detection continues
until an End Peak-based or a new
Time-based section is
encounterd. |
|
Parameter (Time-based
Heart-cut and Peak-based) |
The parameters of the cut depend on the
Function selected.. Click in the cell to
change the parameters:
MHC/HiRes
Click the down-arrow to switch between MHC and
HiRes.
Cut
size The length of the cut in
seconds. The Cut size cannot be changed in
MHC mode.
Cut(s)
The number of cuts. The number of cuts cannot be changed
in MHC mode.
Loop Filling
(Time-based Heart Cut)
The loop filling factor. This value is read-only and cannot be
changed. Multi-Inject
This check box is enabled when the number of cuts is
more than one, Mark the check box to delay the analysis of multiple
cuts until all cuts have been parked.
Analyze
Mode Click the down-arrow and
choose a mode from the drop-down list: Default: the cut is analyzed as soon as
possible. Delayed: analysis is
delayed until there is an available time slot. Ignored: the cut is not
analyzed. Prioritize cut(s) (Time-based
Heart Cut) Mark this check box
to specify that analysis of the cut(s) should be given
priority. Reference Index
Click the down-arrow and select an identifier for the peak as a time
reference. This allows you to adjust for a shift in retention times.
In Time-based heart cutting, the
reference peak parameters are disabled until a reference peak has been
in the Peak-based
prameters. Reference Factor (Time-based Heart
Cut) Enter a factor by which all
Expected retention times are multiplied. This allows you to adjust for
differing relative retention times. This parameter is disabled
until a reference peak has been identified by Peak-based
detection. Expected
Time (Peak-based)
Enter the expected elution time of the reference peak in this
field. Reference
Only (Peak-based)
Mark this check box to indicate that this peak is a reference peak that
should not be
analyzed. |
Parameter (Time-based
Comprehensive) | |
You can set up the Sampling Table either manually or interactively using
the Preview. Any changes you make to the
Sampling Table are synchronized with
the Preview. For manual editing, the
following buttons are also available:
Add |
Inserts a line above the currently selected line in the
Sampling
Table. |
Remove |
Deletes the selected line in the Sampling
Table. |
Clear all |
Clears the Sampling
Table. |
Cut |
Deletes the currently selected line in the Sampling Table and places it on the
clipboard. |
Copy |
Copies the currently selected line in the Sampling Table to the
clipboard. |
Paste |
Pastes the line on the clipboard into the Sampling Table according to its
time. |