Online Sample Manager/Hybrid Multisampler Dashboard Panel

Module Graphic

The items in the Online Sample Manager or Hybrid Multisampler graphic have the following meaning and function:

Actuals

The following Online Sample Manager or Hybrid Multisampler actuals are displayed:

Injection Volume

The volume of the last injection (in µL).

Sample Location

The location of the last sample.

Temperature

The temperature of the cooler or thermostat (in ° C). This actual is present only if a cooler or thermostat is installed.

Prime Status

Either the time remaining for the current prime procedure or Idle (no prime currently in progress).

Auto-clean Status

Either the time remaining for the current auto-clean procedure or Idle (no auto-clean currently in progress).

Active Tray

The number of the tray that is currently on the sample workspace.

Context Menu

The context menu contains the following commands:

Control

Displays the sampler's Control dialog box.

Method

Displays the sampler's Method Setup dialog box.

Injector Program

Displays the sampler's Injector Program .

Identify Device

Causes the LED on the front of the module to blink for a few seconds.

Error Method

  • Set Error Method

    Sets the error method to the method that is currently available in the hardware. The error methods are loaded automatically on all modules if an error occurs.

  • Clear Error Method

    Removes the currently loaded error method.

Home All

Cleans up all pending instructions and moves the device into a default state.

Reset Injector

Sends a reset signal to the sampler. During the reset, the injector is in a Not Ready state (yellow).

Switch Valve to Bypass

Switches the valve from the main flow path to bypass.

Switch Valve to Mainpass

Switches the valve from bypass to the main flow path.

Switch on/off Tray Illumination

Switches the tray illumination on or off.

Auto-clean

Displays the Auto-clean Settings and Start dialog box.

Prime Solvents

Displays the Prime Settings and Start dialog box.

Modify

These parameters override the Configuration settings.

Assign Wellplates

Displays the Tray and Plate Configuration dialog box.