These messages apply to the LC/MSD iQ Single Quadrupole instrument.
If any of these errors continues to occur, copy the error on the screen and contact Agilent for more help.
Carrier Board L1 Fault: Fragmentor fault - the Lens Driver cannot achieve the requested voltage.
Carrier Board L2 Fault: Skimmer fault - the Lens Driver cannot achieve the requested voltage.
Carrier Board L3 Fault: Cyclone Outer DC fault - the Lens Driver cannot achieve the requested voltage.
Carrier Board L4 Fault: Cyclone Exit DC fault - the Lens Driver cannot achieve the requested voltage.
Carrier Board L5 Fault: Cyclone Exit Lens fault - the Lens Driver cannot achieve the requested voltage.
Carrier Board L6 Fault: MS Entrance Lens fault - the Lens Driver cannot achieve the requested voltage.
Carrier Board L7 Fault: MS Shroud DC fault - the Lens Driver cannot achieve the requested voltage.
Carrier Board L8 Fault: Iris Lens fault - the Lens Driver cannot achieve the requested voltage.
The Cyclone Driver cannot maintain the requested inner RF peak voltage.
The Cyclone Driver cannot maintain the requested outer RF peak voltage.
MS U- driver to the Quad cannot maintain the set voltage
MS U+ driver to the Quad cannot maintain the set voltage
MS RF driver to the Quad cannot maintain the set voltage
U+/U- DC to DC converter supply (MS1) is not producing the set voltage
Carrier Board Lens driver power supply cannot maintain the voltage.
The Vertical Interconnect Board is not connected to Carrier Board via the Interconnect Board (ICB).
The Cyclone Driver board is not connected to the Carrier Board via the Interconnect Board (ICB).
"The ribbon cable between the MS1 Quad Driver and the ICB is not connected.
"The Nozzle HV PCA is not connected to the Carrier Board. Not used on Aston 1.
The GFM PCA is not connected to the Carrier Board.
HED Power Supply is not connected to Carrier Board.
The power and lens cable between the MS Quad Driver and the ICB is not connected.
Hardware Shutdown event is coming the Carrier Board. Any of the faults on Status1 and Status2 will cause an Internal Shutdown.
Turbo Fan locked
+48V is out of operational limits on the Interconnect Board.
+24V is out of operational limits on the Interconnect Board.
+15V is out of operational limits on the Interconnect Board.
-15V is out of operational limits on the Interconnect Board.
Interconnect Board CPU Temperature above operating limit.
MS Quad Heater Temperature Sensor reading is out of operational limits.
MS Quad Heater Temperature Sensor failed to return value or sensor harness disconnected or broken.
MS Quad Heater shutoff due to temperature exceeding maximum limit.
Short circuit detected on MS Quad Heater.
+48V_RF is out of operational limits on the Interconnect Board.
Unknown source type
Device driver software error.
Zero Crossing Signal from AC Mains not available.
Q1 Fan Fault
Q2 Fan Fault
+3.3V_ZDB is out of operational limits on the Interconnect Board.
+1.8V_ZDB is out of operational limits on the Interconnect Board.
Current Draw on +48V supply is out of operational limits on the Interconnect Board.
Current Draw on +48V_RF supply is out of operational limits on the Interconnect Board.
Current Draw on +24V supply is out of operational limits on the Interconnect Board.
Current Draw on +15V supply is out of operational limits on the Interconnect Board.
Current Draw on -15V supply is out of operational limits on the Interconnect Board.
Current Draw on +5V supply is out of operational limits on the Interconnect Board.
Current Draw on +5V_ZDB supply is out of operational limits on the Interconnect Board.
Current Draw on +3.3V_ZDB supply is out of operational limits on the Interconnect Board.
Current Draw on +1.8V_ZDB supply is out of operational limits on the Interconnect Board.
+48V is out of operational limits at the output of the Ultimod Power Supply.
+48V_RF is out of operational limits at the output of the Ultimod Power Supply.
+24V is out of operational limits at the output of the Ultimod Power Supply.
+15V is out of operational limits at the output of the Ultimod Power Supply.
+5V is out of operational limits at the output of the Ultimod Power Supply.
Current Draw on +48V supply is out of operational limits at the output of the Ultimod Power Supply.
Current Draw on +48V_RF supply is out of operational limits at the output of the Ultimod Power Supply.
Current Draw on +24V supply is out of operational limits at the output of the Ultimod Power Supply.
Current Draw on +15V supply is out of operational limits at the output of the Ultimod Power Supply.
Current Draw on +5V supply is out of operational limits at the output of the Ultimod Power Supply.
High Vacuum Gauge Filament current is out of operational limits on the Interconnect Board.
AC Mains RMS voltage is out of operational limits.
MS Quad Heater Current is out of operational limits on the Interconnect Board.
Ambient Temperature is out of operational limits on the Interconnect Board.
Humidity inside instrument is out of operational limits.
Current draw on the +48V supply on the Cyclone Driver PCA is out of operational limits.
+48V supply on the Cyclone Driver PCA is out of operational limits.
+5.5V supply on the Cyclone Driver PCA is out of operational limits.
+40V supply on the Cyclone Driver PCA is out of operational limits.
Cyclone Inner Rods driver integrator output is out of operational limits.
+5VA Inner Rods supply on the Cyclone Driver PCA is out of operational limits
Current Monitoring of the +40V for the inner rods output state on the Cyclone Driver is out of operational limits.
+5VA Outer Rods supply on the Cyclone Driver PCA is out of operational limits
Cyclone Outer Rods driver integrator output is out of operational limits.
Current Monitoring of the +40V for the outer rods output state on the Cyclone Driver is out of operational limits.
High Vacuum Ion Gauge Filament Driver in fault state.