Edit MS Global Parameters (LC/MS)
In the Method Editor, you set global MS method parameters that are used when you acquire data. The MS Global Parameters pane is the top pane in the Instrument Setup > SQ > Method section. It is visible when any of the Instrument Setup > SQ > Method sections are visible.
You select the Ion source, the Stop time, and more.
- If necessary, show the MS Global Parameters pane.
-
- Click Method.

-
- In the Acquisition Method window in the navigation pane, click Instrument Setup > SQ.
- In the navigation pane under Method, select any section.
- (optional) Select the Ion Source. The installed Ion Source is displayed next to the selection. You can edit and save a method with an ion source other than the installed source. However, you cannot download or run a method if the sources do not match. If you try to download or run the method, the Not Ready Text in the Instrument Status > SQ tab says "Ion source mismatch."
- Select or clear the Time filter window (min) check box. If you select this check box, then you also enter the peakwidth.
- Click the Stop time. You can click either As pump/No limit or Limit (min).
- If you click Limit (min), then you also enter the time in minutes.
- Select the Tune file. If you select a Fast Scan tune file, then the Fast scan check box is selected automatically.
See Also
Edit a method (LC/MS)
Method > MS Global Parameters Pane (LC/MS)