Reset Spray Chamber parameters (LC/MS)
If possible, get the values from the Autotune report that was shipped with the instrument listing its individual optimum values. If that is not available, use the following defaults.
Electrospray Spray Chamber Parameters
Parameter |
Positive |
Negative |
Drying Gas Temperature (°C) |
300 °C |
300 °C |
Drying Gas Flow (L/min) |
7.0 L/min |
6.0 L/min |
Nebulizer Pressure (psi) |
20 psi |
20 psi |
VCap (V) |
4000 V |
4000 V |
APCI Spray Chamber Parameters
Parameter |
Positive |
Negative |
Vaporizer (°C) |
325 °C |
325 °C |
Corona (µA) |
4 µA |
4 µA |
Drying Gas Temperature (°C) |
300 °C |
300 °C |
Drying Gas Flow (L/min) |
6.0 L/min |
3.0 L/min |
Nebulizer Pressure (psi) |
15 psi |
15 psi |
VCap (V) |
4000 V |
2500 V |
APPI Spray Chamber Parameters
Parameter |
Positive |
Negative |
Vaporizer (°C) |
300 °C |
300 °C |
Lamp |
on |
on |
Gas Temperature (°C) |
350 °C |
350 °C |
Gas Flow (L/min) |
5.0 L/min |
5.0 L/min |
Nebulizer Pressure (psi) |
15 psi |
15 psi |
VCap (V) |
3000 V |
3000 V |
MMI-ES+APCI Spray Chamber Parameters
Parameter |
Positive |
Negative |
Vaporizer (°C) |
200 °C |
200 °C |
Gas Temperature (°C) |
300 °C |
300 °C |
Gas Flow (L/min) |
5 L/min |
5 L/min |
Nebulizer Pressure (psi) |
40 psi |
40 psi |
VCap (V) |
2000 V |
2000 V |
Corona (µA) |
1 µA |
1 µA |
Charging Voltage (V) |
2000 V |
2000 V |
MMI-ES Spray Chamber Parameters
Parameter |
Positive |
Negative |
Vaporizer (°C) |
150 °C |
150 °C |
Gas Temperature (°C) |
250 °C |
250 °C |
Gas Flow (L/min) |
5 L/min |
5 L/min |
Nebulizer Pressure (psi) |
60 psi |
60 psi |
VCap (V) |
2000 V |
2000 V |
Corona (µA) |
N/A |
N/A |
Charging Voltage (V) |
2000 V |
2000 V |
MMI-APCI Spray Chamber Parameters
Parameter |
Positive |
Negative |
Vaporizer (°C) |
250 °C |
250 °C |
Gas Temperature (°C) |
350 °C |
350 °C |
Gas Flow (L/min) |
5 L/min |
5 L/min |
Nebulizer (psi) |
20 psi |
20 psi |
VCap (V) |
2000 V |
2000 V |
Corona (µA) |
5 µA |
5 µA |
Charging Voltage (V) |
2000 V |
2000 V |
AJS ESI Spray Chamber Parameters
Parameter |
Positive |
Negative |
Gas Temperature (°C) |
300 °C |
300 °C |
Gas Flow (l/min) |
5 L/min |
5 L/min |
Nebulizer (psi) |
45 psi |
45 psi |
Sheath Gas Temp (°C) |
250 °C |
250 °C |
Sheath Gas Flow (L/min) |
11 L/min |
11 L/min |
Capillary Voltage (V) |
3500 V |
3500 V |
Nozzle Voltage (V) |
500 V |
500 V |
- Click Method.

- In the Acquisition Method window in the navigation pane, click Instrument Setup > SQ.
- If needed, click Take. In this image, Take is already selected.

- In the navigation pane select Tune > Manual Tune.
- In the navigation pane select Tune > Manual Tune > Source.
|
|
The source parameters are not saved with the tune file. They can be changed and downloaded by you to do some explicit manual tune experimentation, but these parameters will not be saved. The source parameters set in the method are saved and are what is used in acquiring a sample. |
- (optional) Open a different tune file. Click
to open the Load Tune File dialog box. Select a tune file and click Load.
- Click
to lock the instrument for tuning. You need to unlock the instrument after you finish tuning.
- If available, enter the Vaporizer. Refer to the tables above if you cannot find the paper shipped with the software.
- Enter Gas Temp (°C).
- Enter Gas Flow (L/min).
- Enter Nebulizer (psi).
- Enter Capillary Voltage (V).
- If available, enter the Sheath Gas Temp (°C).
- If available, enter the Sheath Gas Flow (l/min).
- If available, enter the Corona (µA).
- If available, enter the Charging Voltage (V).
- Save the tune file.
- Click
to download these parameters to the instrument.
- Click
to unlock the instrument. You need to unlock the instrument after you finish tuning.