The snippet Calibration curve with LOD/LOQ includes the calculated the Limit of detection (LOD) and Limit of quantitation (LOQ), based on the standard deviation of the quantification residuals (Analytical Biochemistry 385 (2009) 101-106). The calculated values may be higher than the usual approach (injection at low concentration to reach signal-to-noise 3.3 or 10) as it evaluates the quantification process in addition to the instrument capability.
The snippet can be used for Single Sequence Summary or Cross-Sequence Summary reporting (see Report Types), and will show the calibration standard results as well as the calibration used for the sample quantification.
Calibration standard results
For all quantified samples other than standard, the snippet will report the calibration curve used for the compound quantification:
Samples other than standard