Configure the 35900E interface
|
|
Proper operation of the 35900E requires that
DIP switches 3, 4, 5, and 6 on the back of the 35900E must be
in the down position. |
The procedure below describes
how to configure a 35900E. You can configure the 35900E to provide its
channel A and channel B resources to one instrument, or you can configure
a single 35900E device in the data system (using the same IP address)
as two instruments and assign channel A resources to one instrument and
its channel B resources to the other instrument.
- If the instrument is not already available in the Control Panel,
add it and select Configure Instrument.
Refer to the data
system help for details.
- Enter a name for this 35900E interface in the Device
Name field.
- Set the IP address. See your LAN administrator for help in assigning
IP addresses. The 35900 requires a Bootp server to assign it an IP
address.
- Enter the instrument IP address in the IP
Address field.
- Click Get Serial Number and Firmware
to connect to the instrument. Upon successful connection, the
instrument serial number and firmware revision appear in the appropriate
fields. If a communication error occurs, check that the 35900E
is connected to the LAN and that it received an IP address from
a BootP server.
- Select the appropriate External Start/Stop option to define
the synchronization at run time between the 35900E, external devices,
and the data system. See 35900E
Interface Configuration for more details.
- Configure how the 35900 front panel buttons will function when
pressed. See 35900E
Interface Configuration for more details. The settings apply to
both channels.
- Select Use this Channel to assign Channel A, Channel B,
or both channels to the interface assigned to this instrument.
- Enter a name in the Channel name field for each activated
channel. This name identifies the attached signal in the data system
and reports.
- Enter units and a conversion multiplier in the Y-Axis Units
and the Y-Axis Multiplier fields for each activated channel.
See Y-Axis
Multiplier fields in 35900E Instrument Configuration for
more details. Note that the Y-Axis Units field is a label for
convenience—it does not change data.
- Click Events to display the Event
Configuration Display dialog. For each pin assigned to an event,
select the check box and enter the required descriptive labels for
the events high and low states. These labels display in the drop-down
list used for assigning events in the method’s event table.
- Click OK when done.
See also:
35900E Instrument
Configuration
Event Configuration Display