The Auto-fill trigger list has to be configured only if using the Automatic command fill-in feature.
Please refer to Automatic filling of analysis parameters to learn more.
The parameters defined as trigger in the Commands window/Auto-fill trigger column, are used by the Sample Scheduler as a ‘key set’ allowing it to select which command template (set of parameters relative to one analysis) must be used to complete an incomplete command generated by the LIMS.
The parameters belonging to the Auto-fill trigger list must mandatorily be assigned a value to allow a command or an analysis to be accepted as template.
The templates are displayed in the Auto-Fills table.
If no trigger is selected, the auto-fill mechanism is not applied. Commands generated by the LIMS are not completed with template. |
If changes are applied to the Sample Scheduler configuration that require a restart of the Sample Scheduler client, a pop-up message appears in the client informing the user that the client needs to be restarted. |