Sample types are used in a sequence
to describe each item being processed. You select sample types in the
sequence table or as part of a sequence template.
Sample type descriptions:
Sample– A sample with unknown
amounts of analytes being analyzed.
Blank– A sample without any
analytes that is treated like a sample. It is used as a signal-to-noise
reference for all subsequent samples and for system suitability. If
an ISTD method is being used, then the Blank usually contains the
internal standard.
Double
blank– Sometimes
called "solvent blank", it is used to prove that no chromatographic
artifacts are coming from the instrument flow path for that particular
column.
Cal. Std. (Calibration
standard) – A sample with a known amount of analyte that
is used as a reference to create or update a calibration table in
the processing method.
QC check– A sample with a known
amount of analyte that is used to verify and prove that the calibration
is correct.
Spike - A sample
with a known amount of standard added.
Sys Suit - A
sample that is run to demonstrate that system is working correctly
as expected. This sample is typically run before running other samples.