Processing method: Extraction: Chromatogram (MS tab)

Here, you set the parameters used to generate Extracted Ion Chromatograms (EICs).

Single m/z expansion chromatograms

This parameter is used for creating EICs from a Total Ion Chromatogram (TIC) or from an MS spectrum.

It is only used for single m/z values, not for mass ranges!

EIC extraction retention time window for target compounds

During quantification or MS sample purity calculation, EICs for each compound are automatically extracted. With the extraction retention time window, you limit the time range around a compound's retention time for which these EICs are extracted. You can thus make sure that the correct peaks are used.

If you provide both absolute and relative time ranges, a sum of both will be used (same principle as for the peak identification window).