Configure ramp and plot ion abundance vs a ramped parameter
This procedure plots the abundance of the 3 masses specified in
the Acquisition Profile Window UI
vs a range of values for a specified analyzer component. A workflow
might include replacing the value in the Parameters
area with the newly determined value. This modification would then
be evaluated by running a scan in manual
tune to determine if this change was acceptable.
- Request tune control and
select Manual Tune > Parameters.
- Select Ramp from the
Acquisition type dropdown
list.
- In the Ramp area, select
a parameter to ramp from the Type dropdown list.
- If necessary, edit the default
Start
and Stop
value range and the Step size for this parameter.
- If needed, select Smooth
ramp data and select the Type of result to display.
- Click Start acquiring data
to start the ramp
and display the plot during the run.
When the data acquisition is complete, a vertical selection bar
displays on the graph.
- Drag the vertical selection bar in the plot to an optimal value
for your ramped parameter and click. This value will display in
the parameter box next to Apply.
- Click Apply to change
the selected parameter to this value.
- Click Save tune parameters
with new tune file name and name this manual tune file.
- Optionally, open a pdf of the manual tune file parameters.
- Release
tune control.