Optimize Lens2 RF parameters during manual tune (LC/MS)

The Lens2 RF is not equipped on the 6120C instrument.

Before you tune the Lens2 RF parameter, you tune the instrument with the Lens2RF Enabled check box cleared. It is important to know that the Lens2RF parameters do affect the resolution of the mass filter. In general, the resolution will improve for all masses if properly tuned, with the effect most noticeable at the higher masses and less so at the lower masses.

  1. In the navigation pane, select Tune > Manual Tune.
  2. In the navigation pane, select Tune > Manual Tune > Optics.
  3. If needed, click Take. In this image, Take is already selected.

  1. If necessary, click to lock the instrument for tuning. You need to unlock the instrument after you finish tuning.
  2. Select the Lens2 RF Enabled check box.
  3. Select Lens2 RF for the Ramp Parameter.
  4. Click the Ramp button.
  5. Select the Lens2 RF value to use. Click Apply.

See Also

Manually tune the MS (LC/MS)

Tune the MS automatically (LC/MS)

Tune the MS using Fast Scan (LC/MS)

Tune > Autotune/Checktune Section