Optimize Lens2 RF parameters during manual tune (LC/MS)
The Lens2 RF is not equipped on the 6120C instrument.
Before you tune the Lens2 RF parameter, you tune the instrument with
the Lens2RF Enabled check box cleared. It is important to know that the
Lens2RF parameters do affect the resolution of the mass filter. In general,
the resolution will improve for all masses if properly tuned, with the
effect most noticeable at the higher masses and less so at the lower masses.
- In the navigation pane, select Tune
> Manual Tune.
- In the navigation pane, select Tune
> Manual Tune > Optics.
- If needed, click Take.
In this image, Take is already selected.

- If necessary, click
to
lock the instrument for tuning. You need to unlock the instrument
after you finish tuning.
- Select the Lens2 RF Enabled
check box.
- Select Lens2 RF for the
Ramp Parameter.
- Click the Ramp button.
- Select the Lens2 RF value
to use. Click Apply.