Prepare for manual tuning (LC/MS)
Preparation for manual tuning consists of loading the atunes.tune file,
making sure the Single Quadrupole LC/MS is set for the correct Ion
Source and Polarity (Positive or Negative),
and making sure the parameters that control ion formation are at acceptable
settings. When tuning with a MultiMode source (MMI), tune in MMI-APCI
mode.
- Click Method.

- In the Acquisition Method window in the navigation pane, click Instrument Setup > SQ.
- In the navigation pane, double-click Tune
> Manual Tune.
- If needed, click Take.
In this image, Take is already selected.

- In the navigation pane, select Tune
> Manual Tune > Source.
- (optional) Open a different tune file. Click
to open the Load Tune File
dialog box. Select a tune file and click Load.
- Click
to lock the instrument for tuning. You
need to unlock the instrument after you finish tuning.
- If the source is MMI, then select the Ion
mode. You can select ESI,
APCI, or ESI+APCI.
- Turn the Calibrant on. Select the Calibrant
on check box. Use the appropriate calibrant.
- Click the Polarity. You
can click Positive or
Negative.
- The Source Parameters change depending on which source is installed.
The default source parameters are loaded for the installed source.
These values are good starting values:
|
|
Because of slight variations from spray chamber
to spray chamber, changes such as adjustment of the nebulizer,
and normal aging of LC/SQ components, these default values may
not be optimum for a particular spray chamber. You can refer to
your Acquisition method for other starting values. |
ESI
Parameter |
Value |
Gas Temperature |
300 (°C) |
Gas Flow |
7 L/minute |
Nebulizer |
15 psi |
Capillary Voltage |
4000 V |
APCI
Parameter |
Value |
Gas Temperature |
300 °C |
Vaporizer |
325 °C |
Gas Flow |
5 L/minute |
Nebulizer |
15 psi |
Capillary Voltage |
4000 V |
Corona Current |
4 µA |
APPI
Parameter |
Value |
Gas Temperature |
300 °C |
Vaporizer |
325 °C |
Gas Flow |
5 L/minute |
Nebulizer |
15 psi |
Capillary Voltage |
4000 V |
MMI-ESI
Parameter |
Value |
Gas Temperature |
350 °C |
Vaporizer |
200 °C |
Gas Flow |
5 L/minute |
Nebulizer |
20 psi |
Capillary Voltage |
2500 V |
Corona Current |
6 µA |
MMI-APCI with APCI Tuning Calibrant
Parameter |
Value |
Gas Temperature |
350 °C |
Vaporizer |
200 °C |
Gas Flow |
5 L/minute |
Nebulizer |
15 psi |
Capillary Voltage |
3000 V |
Corona Current |
4 µA |
Charging Voltage |
2000 V |
AJS ESI
Parameter |
Value |
Gas Temperature |
300 °C |
Gas Flow |
5 L/minute |
Nebulizer |
20 psi |
Sheath Gas Temp |
250 °C |
Sheath Gas Flow |
10 L/minute |
Capillary Voltage |
3500 V in positive mode
3500 V in negative mode |
Nozzle Voltage |
2000 V |
See Also
Manually tune the MS (LC/MS)
Tune the MS automatically
(LC/MS)
Tune the MS using
Fast Scan (LC/MS)
Tune > Autotune/Checktune
Section