Ramp the lenses during manual tune (LC/MS)

In this step, several parameters are ramped to find their optimum settings.

  1. In the navigation pane select Tune > Manual Tune
  2. In the navigation pane select Tune > Manual Tune > MS.
  3. If needed, click Take. In this image, Take is already selected.

  1. If necessary, click to lock the instrument for tuning. You need to unlock the instrument after you finish tuning.
  2. Mark the check boxes next to the masses to ramp.

Positive Polarity

  Electrospray APCI/MMI APPI AJS ESI
Low mass 118.09 121.05 121.05 118.09
Middle mass 922.01 922.01 922.01 922.01
High mass for 6130C & 6135C 2121.90 2121.93 2121.93 2121.90
High mass for 6120C & 6125C 1521.97 1521.97 1521.97 1521.97

Negative Polarity

  Electrospray APCI/MMI APPI AJS ESI
Low mass 112.99 119.04 119.04 112.99
Middle mass 1033.99 805.99 805.99 1033.99
High mass on 6130C & 6135C 2233.90 1805.92 1805.92 2233.90
High mass on 6120C & 6125C 1633.95 1633.95 1633.95 1633.95
  1. In the navigation pane under Tune > Manual Tune, select Optics.
  2. Select Lens1 DC for the Ramp Parameter.
  3. For the ramp parameters, enter -5 for From. Enter 5 for to. Enter 0.2 for the step.
  4. Click Ramp. The software will ramp lens 1 for the selected masses and produce a graph of voltage versus abundance for each mass.
  5. Click the point where abundance is maximized for the highest mass. You get better performance across the entire mass range. Click Apply.
  6. Select Iris for the Ramp Parameter.
  7. For the ramp parameters, enter -400 for From. Enter 400 for to. Enter 5 for the step.
  8. Click Ramp. The software will ramp the Iris for the selected masses and produce a graph of voltage versus abundance for each mass.
  9. Click the point where abundance is maximized for the highest mass. You get better performance across the entire mass range. Click Apply. Typical voltages are 350 for Positive polarity and 0 for Negative polarity.

See Also

Manually tune the MS (LC/MS)

Tune the MS automatically (LC/MS)

Tune the MS using Fast Scan (LC/MS)

Tune > Manual Tune Section (LC/MS)