Be sure both instruments are turned on and connected to the same network as your OpenLAB Control Panel.
In the Control Panel, with your instrument selected (My GCMS in this example), click Configure Instrument.
Under Available modules, double-click the components included in the instrument: Agilent 597x MSD, and Agilent 7890 GC (for example). This moves both to the configured modules column.
In the Configured modules column, double-click each module, one at a time to display the configuration box for that module. See to configure the MSD, and to configure the GC.
The configuration box for the Agilent 597x MSD is similar to the one below.
Enter the IP Address for this instrument in the highlighted box.
Click Get MS Configuration. See Update firmware for details on updating your firmware, if necessary.
Verify that the column is configured to the correct detector. (GC Front Inlet is selected in this example.) If your MS is not configured with a GC, be sure to select Other for this box.
Select the ion source installed in your instrument. (EI, CI, or HES)
Verify the Quad DC Polarity, as shown on the analyzer side door, is correct for this source.
If your instrument has a gas controller, click Gas Control Configuration to define your controller type and gas configuration.
If your instrument does not have a gas controller, skip this step.
The Gas Control Configuration dialog you see will be based on the source you identified.
For details on configuring your specific ion source, see one of the following:
Configure a CI source with JetClean (597x)
Configure an HES or EI source with JetClean (597x)
Configure an HES or EI source without JetClean (597x)
After completing the Gas Control Configuration, you are returned to here once again. Click OK to complete the configuration.
For the Prep Run on Manual Request - Normally, the data system sends a Prep Run command at the start of every run. This command makes sure the GC is prepared before starting an analysis.