Open the Run Queue.
Click Stop .
Select:
Abort current run - aborts the acquiring run. The next run in the sequence is started. This option is not available for a Headspace sampler or for full sequence download runs.
Abort current run and abort sequence - aborts the acquiring run and all runs in the sequence. The next run in the queue is started.
Complete current run and abort sequence - allows the acquiring run to complete but aborts all remaining runs in the sequence. The next run in the queue is started. This option is not available for a Headspace sampler or for full sequence download runs.
Abort all items submitted by current user - aborts all runs in the queue submitted by the logged on user. Runs submitted by other users continue.
Abort all - stops all runs in the queue submitted by any user.
Aborted sequences or single-sample runs remain in the Run Queue with an Aborted status. When a run is aborted, the next valid run is automatically started.